Agilent Turbo-V 1001 Controller - NEW
90 Day Warranty
Agilent Turbo-V 1001 Turbomolecular Vacuum Pump Controller - NEW
Agilent Turbo-V 1001 rack controllers are microprocessor-controlled frequency converters with self-diagnostics and protection features that ensure the highest degree of reliability. The compact, ½ rack unit has a multifunction alphanumeric display for pump status and error code analysis. The front panel has a two-line dot matrix LCD display with back lighting, which displays rotational speed, current, power, and bearing temperature. The microprocessor also acts as a pump cycle log, and can display the number of vacuum cycles, the cycle time for the current cycle, and the total operating hours on the pump. Remote operation can be accomplished with logic level contact closures and with optional computer interfaces.
The Turbo-V 1001 Navigator controllers are microprocessor controlled frequency converters. They are fully controllable through PC software, with self-diagnostics and protection features that ensure the highest degree of reliability. They can be mounted onboard, either on the bottom or on the side of the pump, offering outstanding flexibility and simplicity.
Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!
Specifications | Rack | Onboard |
---|---|---|
Input: Voltage Frequency Max power Stand-by power |
100 ÷ 240 Vac (±10 %), 1-phase 50 to 60 Hz 800 VA - |
100 - 240 Vac 50 to 60 Hz 600 VA 30 to 35 W |
Output: Power during pump ramp-up Power normal (pump water cooled) Power normal (pump air cooled) |
450 W maximum 450 W maximum 300 W maximum |
|
Operating temperature | +5°C to + 45°C | |
Storage temperature | -20°C to + 70°C | |
Fuse | 2 x T10 AH (slow blow) 250 V 5x20 fuses | 1 x 6.3 A |
Serial communication | RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector | |
w/ T-Plus software (optional) | w/ Navigator software (optional) | |
Power cable | With European or NEMA plug 3 meters long (optional) | |
Pump cable | 5 meters long fixed cable | - |
Compliance with | EN 61010 - 1 (2001) EN 61326-1 (2006) Emission level : Class B for Residential Application Immunity level : for Industrial Application |
UNI EN 292-1 UNI EN 292-2 EN-CENELEC 55011 IEC 1000-4-2 (ex 801-2) IEC 1000-4-3 (ex 801-3) IEC 1000-4-4 (ex 801-4) EN 61010-1 (IEC 1010-1) EN 1012-2 |
Max altitude | 2,000 m | |
Protection Category | IP 20 | |
Internal use only | ||
Weight | 6.3 Kg (13.9 lbs) | 5.4 Kg (12 lbs) ISO 160 flange: 19 Kg (41.8 lbs) ISO 200 flange: 19.4 Kg (43 lbs) ISO 250 flange: 21.2 Kg (46.6 lbs) CF 10“ flange: 25.5 Kg (54.2 lbs) |
Installation Category | II | |
Pollution degree | 2 |
Digital Manual - Rack controller
Digital Manual - Onboard controller
Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs. Please follow the above link, and fill out the quote form.
Agilent Turbo-V 1001 Turbomolecular Vacuum Pump Controller - NEW Rack controller X350164003 / X3501-64003 Onboard controller 9698978 / 969-8978