Pumping speed | |
N2 | 80 l/s |
He | 50 l/s |
H2 | 40 l/s |
Compression ratio | |
N2 | 8 x 107 |
He | 2,500 |
H2 | 300 |
Ultimate pressure (measured w/ 2 stages RP & metal seal at inlet) | ≤ 5 x 10-9 mbar |
Maximum pressure in continuous operation | |
TMP cooled by natural convection | 1 x 10-3 mbar at inlet 1 x 10-2 mbar at exhaust |
TMP air cooled | 1 x 10-1 mbar at inlet & exhaust |
TMP water cooled | |
Roughing pump recommended | 4 m3/h (3.2 cfm) - 2004A |
Rotational speed | 27,000 rpm |
Startup | ≤ 1 mn 45 s |
Ambient temperature | |
TMP cooled by natural convection | 0 to 35°C (32 to 95°F) |
TMP air cooled or by natural convection | 0 to 35°C (32 to 95°F) |
TMP water cooled | 0 to 50°C (32 to 132°F) |
Water flow | 0.2 to 1 l/mn w/ water temperature between 10 and 20°C |
Weight | 3 kg (6.6 lbs) |
Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs. Please follow the above link, and fill out the quote form.
Alcatel Annecy Ty. PTM 5081 Turbomolecular Vacuum Pump - REBUILT 795576 795412 5081 ISO-63 115V Ambient air cooled 795574 795412 5081 ISO-63 220V Ambient air cooled 795577 795412 5081 ISO-63 240V Ambient air cooled 795575 795412 5081 ISO-63 100/200V Ambient air cooled 795416 795414 5081 ISO-63 115V Forced air cooled 795413 795519 5081 ISO-63 220V Forced air cooled 795435 795414 5081 ISO-63 240V Forced air cooled 795415 795414 5081 ISO-63 100/200V Forced air cooled 795527 795525 5081 ISO-63 115V water cooled 795529 795525 5081 ISO-63 220V water cooled 795526 795525 5081 ISO-63 240V water cooled 795528 795525 5081 ISO-63 100/200V water cooled