Home > Vacuum Pumps > Cryo > CTI-Cryogenics > Compressors - REBUILT > Brooks CTI-Cryogenics 8510 Compressor - REBUILT
Email Link Tweet This Share on Facebook

Brooks CTI-Cryogenics 8510 Compressor - REBUILT

Request for Quote

To request a quote, click the button below and fill out the form.

RFQ

Manufacturer Part Number: 8031400G002
Brooks CTI-Cryogenics 8510 Compressor - REBUILT
Click Image to Enlarge

This compressor has been designed to operate with water having a pH value of 6.0 to 8.0 and a calcium carbonate concentration of less than 75 parts per million (typical municipal drinking water quality). For applications of lower pH or greater hardness, water conditioning may be necessary.

  Low Voltage High Voltage
Cooling Water
Capacity (pump model) 1 On-Board (10, 400, 500) or
3 On-Board (100, 7, 8, 8F)
Volts 50 Hz: 200 V
60 Hz: 208 / 230 V
50 Hz: 380 V
60 Hz: 460 V
Line frequency 50 / 60 Hz
Phase 3
Full load current (Amps) 50 Hz: 15 A
60 Hz: 17 A
50 / 60 Hz: 8 A
Operating voltage range 50 Hz: 190 - 230 V
60 Hz: 198 - 250 V
50 Hz: 342 - 457 V
60 Hz: 395 - 506 V
Maximum phase run current (Amps) 50 Hz: 23 A
60 Hz: 24 A
50 / 60 Hz: 11 A
Power consumption 5 kW, nominal operating power, 6.75 kW max 5.4 kW, nominal operating power, 6.2 kW max
Nominal power factor 0.8: 7.6 kVA max
Compressor input-power cable (customer supplied)     Recommended type SO-4 conductor, 600 V, neoprene jacket and 10-gauge wire
Helium pressure  Static: 195 - 205 psig (1,345 - 1,415 kPa) at 70 to 80°F (21 to 27°C)
Supply: normal operation: 75 - 85 psig (517 - 590 kPa) at operating temperature.
Ambient operating temperature range     50 - 100ºF (10 - 38ºC)
Interface  
    Cold head power receptacle Mates with plugs on cold head power cable.
    Compressor input-power receptacle     Mates with Hubbel twist lock plug # 2723.
    Gas supply connector 1/2 in. Aeroquip self-sealing coupling, male thread.
    Gas return connector 1/2 in. Aeroquip self-sealing coupling, female thread.
    Remote control receptacle (24 VAC) Mates with P4 connector part # MS3106A-2, supplied by CTI. Mates with J5 connector part # MS3106A-20, supplied by CTI.
Adsorber service schedule Replace every 3 years.
Cooling water requirements 70°F water supply, 2.5 gpm, 14 psid (pressure drop: water inlet to water outlet).
The pressure value does not consider discharge back-pressure conditions
100°F (38°C) maximum outlet temperature
40°F (4°C) maximum inlet temperature
100 psig max nominal inlet pressure
Weight 340 lbs / 154 kg

Digital Manual  low voltage

Digital Manual  high voltage

Request for Quote

Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs. Please follow the above link, and fill out the quote form.

Brooks CTI-Cryogenics 8510 Compressor - REBUILT

8031400G002

Product Reviews

There are currently no customer reviews for this product.

Be the first to write a review!
We value your opinion. Please take a moment and share what you like or don't like about a product.