Edwards nEXT300D - NEW
Edwards nEXT300D Turbomolecular Vacuum Pump - NEW
The nEXT300D is a hybrid bearing compound turbo pump available with either an ISO-100 or a CF-100 inlet flange and has a N2 pumping speed of 300 l/s-1. nEXT pumps combine our proven bearing technology (oil lubricated ceramic lower bearing with dry permanent magnetic upper bearing), an improved rotor design with a new molecular drag stage to deliver improved pumping speed and compression ratios, and true user serviceability. They feature 24V to 48 VDC sensor less motors with a built in drive that is fully compatible our range of TAG and TIC controllers. They are available pre-set for 80W or 160W maximum power, the former enabling use with our 100W TIC controllers are also recommended for pumping Argon, but with a longer ramp time to full operating speed. A basic pump system would require a nEXT300D pump, a TAG or TIC controller, either an air or water cooling accessory or a suitable backing pump.
- Value for money - exceptional pumping speeds and compression ratios
- No unplanned downtime - field proven reliability in the most demanding environments
- Easy integration - standard fit in many scientific instruments
- Reliability for peace of mind - over 200,000 turbomolecular pumps installed
- Low cost of ownership - end user serviceable
Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!
Inlet flange | DN 100 ISO-K / DN 100 CF |
Outlet flange | DN 25 NW |
Pumping speed | |
N2 | 300 l/s-1 |
He | 340 l/s-1 |
H2 | 280 l/s-1 |
Compression ratio D / T | |
N2 | >1 × 1011 |
He | 1 x 108 / > 1 x 108 |
H2 | 5 x 105 / 1 x 106 |
Ultimate pressure with RV backing pump ISO / CF | < 6 x 10-8 / < 5 x 10-10 mbar |
Recommended backing pump* | RV12 / nXDS10i |
Vent port | 1/8 inch BSP female |
Purge port | 1/8 inch BSP female |
Max continuous inlet flow (Nitrogen)† | |
Water cooling (water at 15°C, ambient temp at 40°C) | 95 sccm |
Forced air cooled, 35 °C ambient | 115 sccm |
Pump rotational speed | |
Nominal rotational speed | 60,000 rpm |
Standby rotational speed | Variable from 33,000 to 60,000 rpm (42,000 rpm default) |
Programmable power limit settings | Variable from 50 - 200 W (160 W default) |
Start time to 90% speed D / T 160 W | 155 / 175 |
Analogue outputs | Pump rotational speed; Power consumption; Pump temp; Controller temp |
Cooling method | Forced air / water |
Ambient air temp for forced air cooling | 5 - 35°C |
Min cooling water flow rate (water 15°C) | 15 l/h-1 |
Water temp range | 10 - 20°C |
Max inlet flange temp | 80°C |
Operating attitude | Vertical and upright, through to Horizontal +/- 2° |
Noise level at 1 metre | < 45 dB(A) |
Max magnetic field pump can tolerate | 5 mT |
Recommended controller | TIC 200 turbo and instrument controller |
Quiescent electrical power | Typically 20W |
Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs. Please follow the above link, and fill out the quote form.
Edwards MODEL: nEXT300D Turbomolecular Vacuum Pump - NEW CODE: B822-00-101 / B82200101 nEXT300D ISO-100 80W CODE: B822-00-201 / B82200201 nEXT300D CF-100 80W CODE: B822-00-100 / B82200100 nEXT300D ISO-100 160W B822-00-200 / B82200200 nEXT300D CF-100 160W B823-00-100 / B82300100 nEXT300T HC ISO-100 160W B823-00-101 / B82300101 nEXT300T HC ISO-100 80W B823-00-200 / B82300200 nEXT300T HC CF-100 160W B823-00-201 / B82300201 nEXT300T HC CF-100 80W