PTB Sales: 866-332-0500


Rebuilt Varian Diffusion Pumps

M2, M4 & M6

m
ModelRebuilt
M4$950
M6ask

We also carry Varian Diffusion Pump LN2 Traps

ModelM2
0189
M4
0187
M6
0188
Speed (l/s)
Air1758001,500
Hydrogen2101,0001,800
Helium
Operating Range Untrapped, Torr2 x 10-3 to
<5 x 10-8
1 x 10-3 to
<5 x 10-9
Max. Fore pressure, Torr
No load0.550.650.70
Full load0.400.450.60
Max. Throughput
@ 1 x 10-2 Torr-l/s0.651.52.4
Min. Recommended Backing Pump for Max. Throughput
cfm51011
Backstreaming Rate at inlet flange
(mg/cm2/min)1 x 10-31 x 10-4
Warmup Time (min)1012
Cooldown Time (min) w/ Quick Cool Coil where applicable3020
Fluid Charge100 cm3250 cm3400 cm3
Electrical Requirements1 Φ
50/60 Hz
120/240 V
Power (Watts)4501,1901,785
Cooling water gpm @ 60-80°F (15-26)0.10.150.2

Go to Varian for current user manuals.

What you need, when you need it!