Home > Helium Leak Detectors > VIC > VIC MS-40GT Automatic Portable Helium Leak Detector - NEW

VIC MS-40GT Automatic Portable Helium Leak Detector - NEW

Request for Quote

To request a quote, click the button below and fill out the form.

RFQ

Manufacturer Part Number: 0137-910-00
VIC MS-40GT Automatic Portable Helium Leak Detector - NEW
Click Image to Enlarge

VIC Vacuum Instruments Corporation, LLC MS-40GT Automatic Portable Helium Leak Detector - NEW

Fully automatic, easy to use, and highly accurate, these advanced design leak detectors are ideal for a broad range of applications including clean, oil-free testing.

  • Fully automatic operation, tuning and calibration reduces operator training and error
  • Simple one-button start-up brings the system ready to test in less than three minutes
  • VIC’s dual magnetic sector mass spectrometer provides resolving power of 14: Leak detector does not mistake adjacent peaks for helium
  • Three modes of operation in a single unit: Direct, Reverse, and Combination modes for optimum performance
  • Built-in test diagnostic display system error codes and isolate faults for trouble-free maintenance
  • N.I.S.T. traceable temperature-compensated internal calibrator
  • CE units available

Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!

Sensitivity, vacuum 4.0 x 10-11
Sensitivity, sniffing 6.0 x 10-9
Leak measurement range 1.0 x 10-3 to 4.0 x 10-11
Leak ranging Auto / manual
Start-up time  < 3 min.
Response time  < 2 sec. direct, < 3 sec. reverse
Resolution for 4He 14
Mass spectrometer tube 180° deflection; dual-magnetic sector / stainless steel construction
Auto-tuning Yes
Auto-calibration Internal – yes; External - manual
Auto-zeroing Yes
Reject set-point  Yes
Built-in calibrated leak Yes
NIST traceable Yes
Temperature compensated Yes
He scan Yes
Filaments 2 - Tungsten
Maximum operating pressure     7.5, 760 Torr w/ext roughing option
Display Alpha / numeric
Units atm cc/s, mbar*l/s
Background compensation Continuous automatic compensation for system background
Testable masses 3 and 4
Printer Port Yes
RS-232C Yes
Test modes Direct, reverse or combination flow, sniff
Roughing pump External dry 10.6 cfm or 21.2 cfm dry scroll pump
High vacuum pump 61 liters/second turbomolecular high vacuum pump
Vacuum gauging Pirani: Test port & foreline; Ion: High vacuum
Test port connection NW25
Dimensions (W x H x D) 20.5 x 15 x 14 in. (52.7 x 38.1 x 35.6 cm) 
Weight 96 lbs. (43.6 kg)
Mobile cart with 2-shelves, drawer, power strip, casters
Power 115 V / 50-60 Hz / 8 A; 230-240 V / 50-60 Hz / 5 A; 100 V / 50-60 Hz / 8 A 

Request for Quote

Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs. Please follow the above link, and fill out the quote form.

VIC Vacuum Instruments Corporation, LLC MS-40GT Automatic Portable Helium Leak Detector - NEW

013791000 / 0137-910-00  115V 50-60Hz
013791001 / 0137-910-01  230-240V 50-60Hz
013791002 / 0137-910-02  100V 50-60Hz
013791701 / 0137-917-01  “CE” 230V 50Hz

Clean room

013791300 / 0137-913-00  115V 50-60Hz
013791301 / 0137-913-01  230-240V 50-60Hz
013791302 / 0137-913-02  100V 50-60Hz
013791901 / 0137-919-01  “CE” 230V 50Hz dry