Adixen ATH 400 M - REBUILT
Call for Price
In stock
SKU
REBUILT_ATH400M
| Weight | 42 lbs |
|---|---|
| Pumping Speed | 868.74 cfm |
| Ultimate Pressure | 8 × 10-9 mbar |
Pfeiffer Adixen Alcatel ATH 400 M Turbomolecular Vacuum Pump - REBUILT
One year warranty. We offer Pfeiffer Adixen Alcatel turbo pump repair service. View our inventory of Adixen Alcatel turbo pumps.
Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!
| Inlet flange | 100 ISO-K | 160 ISO-K |
| Rotation speed | 39,000 rpm | |
| Pumping speed | ||
| N2 | 320 l/s | 410 l/s |
| He | 290 l/s | 360 l/s |
| H2 | 180 l/s | 230 l/s |
| Compression rate | ||
| N2 | 1 x 108 | |
| He | 2 x 103 | |
| H2 | 1 x 102 | |
| Ultimate pressure w/out purge measured according to Pneurop standard |
8 x 10-9 mbar | |
| Maximum pressure at inlet in continuous operation | 1 mbar | |
| Maximum permissible pressure at exhaust | 5 mbar | |
| Noise level | ≤ 39 dBA | |
| Start-up time | < 3 min | |
| Maximum start-up power | 650 W | |
| Maximum operating power | 300 W | |
| N2 purge flow rate | 50 SCCM | |
| Cooling water flow rate | < 60 l/h | |
| Water temperature | 15 < T < 25°C | |
| Maximum ambient temperature | 40°C | |
| Weight | 19 kg / 42 lbs | |
| Recommended forepump | ADP 31 | |
| Type | Label | Description | Download |
|---|---|---|---|
| Alcatel ATH 400M / 1000M - REBUILT | Maglev Hybrid Turbomolecular Pumps | Download |
Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs.